3269 Hilton Rd. Ferndale, MI 48220
Certification Accredition By A2LA
Traceable to National Institute of Standard
Equipment Services Group IIc, Calibration Certification Scope of Work
We are fully accredited to verify and calibrate all environmental test chambers. Below and attached is a complete list of all the equipment used in measuring Acoustic and vibration, DC and Low-Frequency Electrical Voltage, Thermocouple Simulation for temperature conditions, and Relative Humidity.
If you click on the “Organization/Acceditation Information” graphics it will take you to our official certificate at A2LA. The official A2LA document of our scope of work is also attached.
The Right Equipment For Measurements
We use the 743B Documenting Process Calibrators as our primary calibrator. This hand-held instrument measures and sources electrical and physical parameters. Our 743B includes the TSLE for thermal couples and PRT for Platinum Resistance Thermometers.
This calibrator lets us troubleshoot, calibrate, verify, and document the work performed and validates the results.
Should you want to review the details specifications on this equipment, click here to download the manual on the Fluke 743B
The Right Pressure Transducer For Measurements
We use the Fluke 700PD7 Dual Pressure Module for all our pressure measurements between -15 and 200 psi. This Dual Pressure Module is directly attached to the Fluke 754B Calibrator to measure the results.
Should you want to review the details on the datasheet for this transducer used, click here to download the manual on the Fluke 743B
The Right Equipment For Measurements
We use the 743B Documenting Process Calibrators as our primary calibrator. This hand-held instrument measures and sources electrical and physical parameters. Our 743B includes the TSLE for thermal couples and PRT for Platinum Resistance Thermometers.
This calibrator lets us troubleshoot, calibrate, verify, and document the work performed and validates the results.
Should you want to review the details specifications on this equipment, click here to download the manual on the Fluke 743B
The Right Pressure Transducer For Measurements
Calibration Certification: Electrical - DC/Low Frequency
Parameter DC Voltage – Measure3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
(0 to 110) mV (0 to 1.1) V (0 to 11) V (0 to 110) V (0 to 300) V |
0.029 % + 20 μV 0.029 % + 0.065 mV 0.029 % + 0.66 mV 0.058 % + 6.4 mV 0.057 % + 21 mV |
Parameter DC Voltage – Generate3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
(0 to 110) mV (0 to 1.1) V (0 to 15) V |
0.011 % + 6.8 μV 0.011 % + 0.065 mV 0.012 % + 0.87 mV |
Parameter DC Current – Generate3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
(2 to 22) mA | 0.013 % + 3.7 μA |
Parameter Resistance – Measure3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
(0 to 110) Ω (0 to 1.1) kΩ (0 to 11) kΩ |
0.058 % + 0.06 Ω 0.058 % + 0.58 Ω 0.12 % + 12 Ω |
Parameter Resistance – Generate3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
(0 to 11) Ω (0 to 110) Ω (0 to 1.1) kΩ (0 to 11) kΩ |
0.011 % + 0.023 Ω 0.011 % + 0.047 Ω 0.023 % + 0.58 Ω 0.03 % + 6.0 Ω |
Parameter Thermocouple Simulation3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
Type J Type K Type T |
Type J Type K Type T |
Parameter RTD Simulation3 | Equipment Fluke 743B |
Range | CMC 2, 4 (+/-) |
100 Ω, Pt (3926) 100 Ω, Pt (385) 100 Ω, Pt (3916) |
100 Ω, Pt (3926) 100 Ω, Pt (385) 100 Ω, Pt (3916) |
Parameter/Equipment | Range | CMC 2, 4 (+/-) | Comments |
---|---|---|---|
DC Voltage – Measure3 | (0 to 110) mV (0 to 1.1) V (0 to 11) V (0 to 110) V (0 to 300) V |
0.029 % + 20 μV 0.029 % + 0.065 mV 0.029 % + 0.66 mV 0.058 % + 6.4 mV 0.057 % + 21 mV |
Fluke 743B |
DC Voltage – Generate3 | (0 to 110) mV (0 to 1.1) V (0 to 15) V |
0.011 % + 6.8 μV 0.011 % + 0.065 mV 0.012 % + 0.87 mV |
Fluke 743B |
DC Current – Generate3 | (2 to 22) mA | 0.013 % + 3.7 μA | Fluke 743B |
Resistance – Measure3 | (0 to 110) Ω (0 to 1.1) kΩ (0 to 11) kΩ |
0.058 % + 0.06 Ω 0.058 % + 0.58 Ω 0.12 % + 12 Ω |
Fluke 743B |
Resistance – Generate3 | (0 to 11) Ω (0 to 110) Ω (0 to 1.1) kΩ (0 to 11) kΩ |
0.011 % + 0.023 Ω 0.011 % + 0.047 Ω 0.023 % + 0.58 Ω 0.03 % + 6.0 Ω |
Fluke 743B |
Thermocouple Simulation3
Type J Type K Type T |
(-200 to -100) ºC (-250 to -200) ºC |
0.55 ºC 1.20 ºC |
Fluke 743B |
RTD Simulation3
100 Ω, Pt (3926) 100 Ω, Pt (385) 100 Ω, Pt (3916) |
(-200 to 0} ºC (-200 to -190) ºC |
0.24 ºC 0.40 ºC |
Fluke 743B |
Calibration Certification: Mechanical
Parameter Pressure3 Gages and Transducers | Equipment Fluke 743B and Fluke 700PD7 Pressure Transducer |
Range | CMC 2, 4 (+/-) |
(0 to 200) psig | 0.19 psig |
Parameter/Equipment | Range | CMC 2, 4 (+/-) | Comments |
---|---|---|---|
Pressure3 Gages and Transducers |
(0 to 200) psig | 0.19 psig | Fluke 743B and Fluke 700PD7 Pressure Transducer |
Calibration Certification:Thermodynamics
Parameter Relative Humidity3 Measure | Equipment Vaisala Temperature and Humidity Indicator |
Range | CMC 2, 4 (+/-) |
(10 to 90) % RH (90.1 to 95) % RH |
1.9 % RH 2.6 % RH |
Parameter Temperature3 Measure | Equipment Fluke 743B w/ Type TSLE TC Fluke 743B w/ PRT |
Range | CMC 2, 4 (+/-) |
(-70 to 200) ºC (-40 to 125) ºC |
1.2 ºC 0.65 ºC |
Parameter/Equipment | Range | CMC 2, 4 (+/-) | Comments |
---|---|---|---|
Relative Humidity3 Measure |
(10 to 90) % RH (90.1 to 95) % RH |
1.9 % RH 2.6 % RH |
Vaisala Temperature and Humidity Indicator |
Temperature – Measure3 |
(-70 to 200) ºC (-40 to 125) ºC |
1.2 ºC 0.65 ºC |
Fluke 743B w/ Type TSLE Fluke 743B w/ PRT |
Calibration Certification: Stipulations
1This laboratory offers commercial calibration service and field calibration service.
2Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.
3Field calibration service is available for this calibration and this laboratory meets A2LA
R104 – General Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer’s site can normally be expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such as the environment at the place of calibration and for other possible adverse effects such as those caused by transportation of the calibration equipment.
The usual allowance for the actual uncertainty introduced by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4The measurands stated are generated with the Fluke 743B instruments. This capability is suitable for the calibration of the devices intended to measure the stated measurand in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a fraction/percentage of the reading plus a fixed floor specification.